doi: 10.17586/2226-1494-2023-23-5-1073-1076
Measurement of the refractive index using an autocollimation goniometer
Read the full article ';
Article in Russian
For citation:
Abstract
For citation:
Yurin A.I., Vishnyakov G.N., Minaev V.L., Krasivskaya M.I. Measurement of the refractive index using an autocollimation goniometer. Scientific and Technical Journal of Information Technologies, Mechanics and Optics, 2023, vol. 23, no. 5, pp. 1073–1076 (in Russian). doi: 10.17586/2226-1494-2023-23-5-1073-1076
Abstract
The paper proposes a modified method for measuring the refractive index of a triangular prism using an autocollimation angle measuring system designed to measure angles formed by flat surfaces of objects. The method involves using a fixed mirror to reflect the refracted beam, measurement of the angles of incidence of the beam on the face of the prism corresponding to the autocollimation positions and calculating the refractive index of the prism material using the solution of a system of equations. The paper presents the experimental study results for a triangular prism made of K8 optical glass using the proposed method, and a comparison of the results with the readings obtained using the State Primary Standard of the Refractive Index Unit GET 138-2021. The proposed method makes it possible to simplify the process of measuring the refractive index since there is no need to measure the angle of deviation of the beam.
Keywords: refractive index, prism method, goniometer, autocollimator, refractometry
References
References
- Refractometric Methods in Physicochemical Measurements. Ed. by L.A. Konopelko. Moscow, Triumf Publ., 2020, 208 p. (in Russian). https://doi.org/10.32986/978-5-907052-08-03-2020-208
- Tilton L.W. Prism Refractometry and Certain Goniometrical Requirements for Precision (Classic Reprint). Forgotten Books, 2018, 24 p.
- Afanasev V.A. Optical Measurements. Moscow, Vysshaja shkola Publ., 1981, 229 p. (in Russian)
- VishnyakovG.N., LevinG.G., KornyshevaS.V., ZyuzevG.N., Lyudomirskiǐ M.B., PavlovP.A., FilatovYu.V. Measuringtherefractive index on a goniometer in the dynamic regime. Journal of Optical Technology, 2005, vol. 72, no. 12, pp. 929–933. https://doi.org/10.1364/JOT.72.000929
- Ioffe B.V. Refractometric Methods in Chemistry. Leningrad, Himija Publ., 1974, 400 p. (in Russian)
- Astrua M., Pisani M. Prism refractive index measurement at INRiM. Measurement Science and Technology, 2009, vol. 20, no. 9, pp. 095305. https://doi.org/10.1088/0957-0233/20/9/095305
- Yurin A.I., Vishnyakov G.N., Minaev V.L. Measurement of the refractive index using a goniometric system. Optics and Spectroscopy, 2022, vol. 130, no. 12, pp. 1624–1627. https://doi.org/10.21883/EOS.2022.12.55252.4103-22
- Yurin A.I., Vishnyakov G.N., Minaev V.L. Measurement of the refractive index using a modified constant deviation method. Measurement Techniques, 2023, vol. 65, no. 1, pp. 904–908. https://doi.org/10.1007/s11018-023-02168-y
- Yurin A.I., Vishnyakov G.N., Minaev V.L. Refractive index measurement using a modified prism method. Measurement Techniques, 2023, vol. 66, no. 2, pp. 96–100. https://doi.org/10.1007/s11018-023-02195-9
- Demchuk V.Iu., Zaitcev I.I. Method for Measuring the Refractive Index of Optical Glass. Patent 1511647, 1987. (in Russian)
- Vishnyakov G.N., Minaev V.L., Bochkareva S.S. GET 138-2021 state primary refractive index standard. Measurement Techniques, 2022, vol. 65, no. 5, pp. 307–314. https://doi.org/10.1007/s11018-022-02090-9